User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Honeywell 8476:Digital Temperature Indicator DATE: 07-Oct-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 14 NUMBER OF LINES: 123 CONFIGURATION: Fluke 5700A STANDARD: Omega TRC III ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P U F W 1.002 ASK+ X 1.003 STD Omega TRC III 1.004 HEAD INITIAL CONDITIONS 1.005 DISP Insert the reference thermocouple into the center 1.005 DISP well of the Ice Point Cell. 1.006 DISP Connect the copper end (white connector) to the 1.006 DISP 5700 OUTPUT HI/LO. Red wire is HI, green wire is LO. 1.007 DISP Connect the thermocouple side (blue connector) to 1.007 DISP channel 1 on the UUT. 1.008 DISP Set the channel selector switch to 1. 1.009 HEAD {* ThermoCouple Tests *} 1.010 HEAD {"T" Type T/C} 1.011 HEAD {Channel 1} 1.012 JMP 2.001 1.013 EVAL 2.001 5700 -2.581mV S 2W 2.002 MEMI Use the keyboard to enter the UUT reading in °F. 2.003 ACC -100°F 0.10U 2.004 MEME 2.005 MEMC °F 3.5U #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 3.001 5700 -0.674mV S 2W 3.002 MEMI Use the keyboard to enter the UUT reading in °F. 3.003 ACC 0°F 0.10U 3.004 MEME 3.005 MEMC °F 3.5U #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 4.001 5700 0.000mV S 2W 4.002 MEMI Use the keyboard to enter the UUT reading in °F. 4.003 ACC 32°F 0.1U 4.004 MEME 4.005 MEMC °F 3.5U #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 5.001 5700 1.518mV S 2W 5.002 MEMI Use the keyboard to enter the UUT reading in °F. 5.003 ACC 100°F 0.1U 5.004 MEME 5.005 MEMC °F 3.5U #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 6.001 5700 3.967mV S 2W 6.002 MEMI Use the keyboard to enter the UUT reading in °F. 6.003 ACC 200°F 0.1U 6.004 MEME 6.005 MEMC °F 3.5U #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 7.001 5700 6.647mV S 2W 7.002 MEMI Use the keyboard to enter the UUT reading in °F. 7.003 ACC 300°F 0.1U 7.004 MEME 7.005 MEMC °F 3.5U #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 8.001 5700 9.523mV S 2W 8.002 MEMI Use the keyboard to enter the UUT reading in °F. 8.003 ACC 400°F 0.1U 8.004 MEME 8.005 MEMC °F 3.5U #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 9.001 5700 10.272mV S 2W 9.002 MEMI Use the keyboard to enter the UUT reading in °F. 9.003 ACC 425°F 0.1U 9.004 MEME 9.005 MEMC °F 3.5U #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 10.001 DISP Move the reference probe thermocouple leads 10.001 DISP to channel 2. Set the channel selector switch to 2. 10.002 5700 1.518mV S 2W 10.003 MEMI Use the keyboard to enter the UUT reading in °F. 10.004 ACC 100°F 0.1U 10.005 MEME 10.006 MEMC °F 3.5U ch2 #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 11.001 DISP Move the reference probe thermocouple leads 11.001 DISP to channel 3. Set the channel selector switch to 3. 11.002 5700 1.518mV S 2W 11.003 MEMI Use the keyboard to enter the UUT reading in °F. 11.004 ACC 100°F 0.1U 11.005 MEME 11.006 MEMC °F 3.5U ch3 #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 12.001 DISP Move the reference probe thermocouple leads 12.001 DISP to channel 4. Set the channel selector switch to 4. 12.002 5700 1.518mV S 2W 12.003 MEMI Use the keyboard to enter the UUT reading in °F. 12.004 ACC 100°F 0.1U 12.005 MEME 12.006 MEMC °F 3.5U ch4 #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 13.001 DISP Move the reference probe thermocouple leads 13.001 DISP to channel 5. Set the channel selector switch to 5. 13.002 5700 1.518mV S 2W 13.003 MEMI Use the keyboard to enter the UUT reading in °F. 13.004 ACC 100°F 0.1U 13.005 MEME 13.006 MEMC °F 3.5U ch5 #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). 14.001 DISP Move the reference probe thermocouple leads 14.001 DISP to channel 6. Set the channel selector switch to 6. 14.002 5700 1.518mV S 2W 14.003 MEMI Use the keyboard to enter the UUT reading in °F. 14.004 ACC 100°F 0.1U 14.005 MEME 14.006 MEMC °F 3.5U ch6 #! Test Tol 3.5, Sys Tol 0.1, TUR 35.0000 (>= 4.00). #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.